Ohaus VX SERIES FLOOR SCALES Indicator T32XW Manual en User Manual
Page 20

EN-
3000 Series Indicators
Reset
No, Yes
Legal for Trade
Off, On
Cal Unt
kg, lb
Capacty
…20000
Graduaton
0.00…20
Power On Unt g, kg, lb, oz, lb:oz,
Auto
Zero Range 0%,
2%, 100%
End Setup Ext SETUP menu
3. Setup Menu
When the Indcator s used for the frst tme, enter ths
menu to set the Capacty and Graduaton.
3.. Reset
Reset the Setup menu to the factory defaults.
No = not reset.
Yes = reset.
NOTE: If the Legal for Trade menu tem s set to ON, the Capacty, Graduaton, Zero Range and Legal For
Trade settngs are not reset.
3.. Capacity
Set the scale capacty from 5 to 20000. Refer to the Setup Table 3.3 for avalable settngs.
3..2 Legal for Trade
Set the legal for trade status.
OFF = off
ON = on
Turnng on the “LFT” menu settng has the followng effects:
• Zero-range s set and locked on “2”.
• Auto Zero Trackng s set and locked on 0.5d
• The lb:oz unt s not avalable as a power-on settng.
3..3 Calibration Unit
Set the unt durng calbraton.
CAL UN kg = Calbrate usng kg weghts
CAL UN lb = Calbrate usng pound weghts