Adventurer

TM

Pro

Analytical and Precision Balances

Added Flexibility of Available Options

The Adventurer Pro also now offers InCal

TM

internal calibration. InCal ensures weighing accu-

racy by allowing you to calibrate the balance prior to use without the need of external weights.


All Adventurer Pro models come with a built-in RS-232 interface, and for added functionality

can be equipped with either a second RS-232 or USB interface, allowing connections to multiple

peripherals. With full GLP/GMP data output, Adventurer Pro can meet traceability and compliance requirements.

QuadraStance

TM

Design for Superior Stability

QuadraStance™ incorporates four adjusting feet to provide superior footprint stability on

any surface. QuadraStance helps prevent rocking and tipping, enhancing the weighing

stability of the balance, and keeping the Adventurer Pro level and accurate.

Additionally, Adventurer Pro has been designed with an up-front level indicator eliminating

the need to look behind the balance during the leveling process. A quick glance at the

up-front level indicator helps to ensure that the balance is level prior to each use.

Other Standard Features Include:

Multiple Application Modes Including:

Animal Weighing Mode – This mode allows the user to weigh small animals and filters

out animal movements to provide a stable weight indication

Parts Counting with Optimization – This mode allows the user to count based on a

calculated piece weight. With optimization, counting accuracy is improved through

auto re-calculation of the piece weight

Percentage Mode – This mode allows the user to display the weight of a sample as a

percentage of a reference weight

Checkweighing Mode – This mode allows the user to check the weight of a sample

against a preset target

Totalization Mode – This mode allows the user to sum a series of sample weights

Display Hold Mode – This mode allows the user to hold the highest stable weight on the display until cleared

Selectable Environmental Settings

Three filter modes and adjustable zero tracking are ideal for working in slow filling applications where sensitivity is

a must. When working in harsh environments, user can adjust balance settings to compensate for vibrations and

other disturbances.

Protective In-Use Cover
The in-use cover is replaceable and offers protection against dust and debris, minimizing the chance of damage to

the display and keypad buttons.