EN-8 SD SERIES BENCH SCALE

Calibration

For best results, calbrate the scale at regular ntervals. (Calbraton weghts are not
suppled wth the scale.)

CAUTION

USE EXTREME CARE WHEN HANDLING CALIBRATION WEIGHTS

AS THEY ARE VERY HEAVY. IMPROPER LIFTING METHODS OR

MISUSE OF CALIBRATION WEIGHTS MAY RESULT IN PERSONAL

INJURY. MULTIPLE WEIGHTS MAY BE USED TO EQUAL THE

REQUIRED CALIBRATION WEIGHT.


1. Allow the scale to warm up and stablze for a least two mnutes before

commencng the calbraton procedure.

2. To select the weghng unt to be used for calbraton, press the

UNITS

Cal

button untl the correct unt s dsplayed.

NOTE: The calbraton process can be stopped at any tme by pressng the

UNITS

Cal

button or by turnng the scale off. The correct span calbraton

weghts wll need to be avalable before startng the calbraton (see

specfcaton table). Calbraton s unavalable n Dynamc Weghng Mode.

3. Press and hold the

UNITS

Cal

button untl “CAL” s dsplayed.

4. Wth the platform empty, press the

ON/ZERO

Off

button to

capture the scale zero. The dsplay wll show “-C-”