VEGA VEGAFLEX 61 Foundation Fieldbus User Manual
Page 14
Advertising
![background image](/manuals/262771/14/background.png)
If possible, avoid sockets. Mount the sensor flush with the vessel top. If
this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used. They simply increase the upper dead band. Check if this is
relevant for your measurement.
In such cases, always carry out a gating out of false signals after
installation. You can find further information under "Setup procedure".
h
d
d
h
DN40 ... DN150
£ 150
> DN150 ... DN200
£ 100
Fig. 6: Mounting socket
When welding the socket, make sure that the socket is flush to the
vessel wall.
1
2
Fig. 7: Socket must be installed flush
1
Unfavourable installation
2
Socket flush - optimum installation
Socket
14
VEGAFLEX
61 • Foundation Fieldbus
4 Mounting
31836
-EN
-100426
Advertising