VEGA VEGAPULS 63 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - four-wire User Manual

Page 59

Advertising
background image

59

11 Supplement

VEGAPULS 63 • 4 … 20 mA/HART - four-wire

36512-EN-130228

3

4

2

1

Fig. 42: Data of the input variable
1 Reference plane

2 Measured variable, max. measuring range

3 Length antenna cone

4 Useful measuring range

Standard electronics
Max. measuring range

35 m (114.8 ft)

Recommended measuring range

Ʋ Flange DN 50, 2"

up to 15 m (49.21 ft)

Ʋ Flange DN 80, 3"

up to 35 m (114.8 ft)

Electronics with increased sensitivity
Max. measuring range

75 m (246.1 ft)

Flange DN 50, 2"

up to 15 m (49.21 ft)

Flange DN 80, 3"

up to 35 m (114.8 ft)

Output variable
Output signals

4 … 20 mA/HART - active; 4 … 20 mA/HART - passive

Terminal voltage passive

9 … 30 V DC

Shortcircuit protection

Available

Potential separation

Available

Signal resolution

0.3 µA

Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current

22 mA

Starting current

≤ 3.6 mA

Load (4 … 20 mA/HART - active)

< 500 Ω

Damping (63 % of the input variable)

0 … 999 s, adjustable

HART output values

Ʋ PV (Primary Value)

Distance to the level

Ʋ SV (Secondary Value)

Level as percentage value

Ʋ TV (Third Value)

Linearised percentage value

Advertising