69

11 Supplement

VEGAFLEX 81 • 4 … 20 mA/HART - four-wire

42280-EN-130314

L

0

-2 mm (-0.079")

2 mm (0.079")

-10 mm (0.394")

5 mm (0.197")

0,15 m (5.906")

0,03 m (1.181")

0,05 m

(1.969")

1

Fig. 31: Deviation VEGAFLEX 81 in coax version in water
1 Dead band - no measurement possible in this area

L Probe length

L

0

-2 mm (-0.079")

2 mm (0.079")

10 mm (0.394")

5 mm (0.197")

-5 mm (-0.197")

0,15 m (5.906")

0,1 m (3.937")

0,05 m

(1.97")

0,1 m

(3.937")

1

1

Fig. 32: Deviation VEGAFLEX 81 in coax version in oil
1 Dead band - no measurement possible in this area

L Probe length

Repeatability

≤ ±1 mm

Variables influencing measurement accuracy
Specifications for the digital measured value
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm (0.394 in)

Additional deviation through electromag-

netic interference acc. to EN 61326

< ±10 mm (< ±0.394 in)

Specifications apply also to the current output
Temperature drift - Current output

±0.03 %/10 K relating to the 16 mA span max. ±0.3 %

Deviation on the current output by ana-

logue/digital conversion

< ±15 µA

Additional deviation through electromag-

netic interference acc. to EN 61326

< ±150 µA