background image

15

4 Mounting

VEGABAR 81 • Slave sensor for electronic differential pressure

45049-EN-131011

4.4 Level measurement

The master/slave combination is suitable for level measurement in a

pressurized vessel
Keep the following in mind when setting up the measuring system:

•

Mount the master sensor below the min. level

•

Do not mount the master sensor close to the filling stream or

emptying area

•

Mount the master sensor so that it is protected against pressure

shocks from the stirrer

•

Mount the slave sensor above the max. level

1

2

Fig. 9: Measurement setup, level measurement in pressurized vessel
1 VEGABAR 81

1 VEGABAR 81 - Slave sensor

4.5 Differential pressure measurement

The master/slave combination is suitable for differential pressure

measurement
Take note of the following instructions for the measurement setup, for

example in gases:

•

Mount the instruments above the measuring point

Possible condensation can then drain off into the process line.

Measurement setup

Measurement setup