3 Product description
VEGASWING 66 • - Relay
44950-EN-130805
3.2 Principle of operation
VEGASWING 66 is a point level sensor with tuning fork for point level
detection.
It is designed for industrial use in all areas of process technology and 
can be used in liquids. It is particularly suitable for applications with
high temperatures up to 450 °C (842 °F) and high process pressure
up to 160 bar (2320 psig).
Typical	applications	are	overfill	and	dry	run	protection.	The	small	
tuning fork allows use in pipelines. Thanks to its simple and rugged
measuring system, VEGASWING 66 is virtually unaffected by the
chemical and physical properties of the liquid.
It	functions	even	under	difficult	conditions	such	as	turbulence,	foam	
generation, buildup, strong external vibration or changing products.
Function monitoring
The electronics module of VEGASWING 66 continuously monitors via
frequency evaluation the following criteria:
•
Strong corrosion or damage on the tuning fork
•
Loss of vibration
•
Line break to the vibrating drive
If a malfunction is detected or in case of power failure, the electron-
ics takes on a defined switching condition, i.e. the relay deenergises
(safe condition).
The tuning fork vibrates at its mechanical resonance frequency of
approx. 1400 Hz. When the tuning fork is submerged in the product,
the frequency changes. This change is detected by the integrated
electronics module and converted into a switching command.
VEGASWING 66 can be operated without external evaluation system.
The integrated electronics evaluates the level signal and outputs a
switching signal. With this switching signal, a connected device can
be operated directly (e.g. a warning system, a pump etc.).
The	data	for	power	supply	are	specified	in	chapter	"Technical data".
3.3 Adjustment
With the factory setting, products with a density > 0.07 g/cm³
(0.025 lbs/in³) can be detected. The instrument can be adapted to
products with lower density.
On	the	electronics	module	you	will	find	the	following	display	and	
adjustment elements:
•
Signal lamp for indication of the operating status (green)
•
Signal lamp for indication of the switching status (yellow)
•
Control lamp for fault indication (red)
•
DIL switch for sensitivity adjustment
•
Mode switch for selecting the switching behaviour (min./max.)
Application area
Functional principle
Voltage supply