Atlas SCR User Guide

January 2008 – Rev 3

Page 10

Gate Sensitivity


The Peak Atlas SCR attempts to trigger the device under test at eight discrete
gate currents, in ascending order.

Trigger test level 1

100

µA

Trigger test level 2

500

µA

Trigger test level 3

2.5mA

Trigger test level 4

10mA

Trigger test level 5

25mA

Trigger test level 6

50mA

Trigger test level 7

75mA

Trigger test level 8

90mA


Thus, for example, a reported trigger current of 10mA means that device
triggering occurred at a gate current of 10mA, but not at 2.5mA. Therefore the
trigger current for the device under test is between 2.5mA and 10mA.

Although Triacs are tested in both quadrants 1 and 3, the reported gate trigger
current (for both SCRs and Triacs) is that which applies to quadrant 3.