Specifications
Operating Specifications
1
Analysis Modes
IC, NPOC, TC, TOC (TC-IC)
Dynamic TOC Range
Up to 50,000 ppm TOC
TOC Limit of Detection (LOD)
0.05 ppm as NPOC
TOC Accuracy**
± 3%
TOC Precision**
RSD ≤ 1.5%
TOC Linearity
R
2
≥ 0.995, measured as NPOC
Analysis Time
2.6 to 9.2 min, depending
on mode
TOC Calibration Stability
Up to 6 months
Particle Diameters in Sample ≤ 800 µm diameter
Ambient Temperature Range 10–40 °C (50–104 °F)
Maximum Relative Humidity
Up to 95%, non-condensing
Sample Temperature Range
10–60 °C (50–140 °F)
Sample Volume 0.6 ml to 3 ml
Sample Inlet Pressure
Ambient
Maximum Backpressure
Gravity drain
Analyzer Specifications
Outputs
Ethernet (1), USB (3)
Display
Color, touch-sensitive LCD
Power Requirements
100–240 ±10% VAC, 400 W, 50/60 Hz
Dimensions
Analyzer: H: 52.05 cm (20.5 in); W: 32.26 cm (12.7 in); D: 58.42 cm (23.0 in)
GE Autosampler: H: 52.2 cm (20.5 in); W: 28.2 cm (11.1 in); D: 53.3 cm (21.0 in)
Weight
Analyzer: 22.41 kg (49.4 lb); GE Autosampler: 14.1 kg (31.1 lbs)
Safety Certifications
CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2 No. 61010-1.
Consumables
Maintenance Frequency
Estimated Maintenance Time
Acid Reagent
As needed, typically for 6 months (285 mL)
5 minutes
Oxidizer Reagent
As needed, typically 30-day stability
5 minutes
Reagent Grade Water
As needed
5 minutes
Air Filter Cartridge
3 months of operation
5 minutes
Pumps
24 months
15 minutes
Sample Flow Path Tubing
6-12 months
15 minutes
Valve Seals (2)
Up to 3 months of operation, depending on sample matrix
15 minutes
* Trademark of General Electric Company; may be registered in one or more countries.
** Measured at 25 ppm NPOC or TOC
*** See Operation & Maintenance Manual for calibration recommendation
1
Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
The Sievers InnovOx TOC Analyzer is protected by one or more of the following US and foreign patents: US 8,101,417, 8,101,418, 8,101,419, 8,101,420
and 8,114,676 .
InnovOx Laboratory Analyzer Details
InnovOx Measurements for Brine,
Cellulose, and Humic Acid Samples
Replicate
28% Brine
Solution
(Process Sample)
90 m Cellulose
Solution
(100 ppm C)
Humic Acid
Solution
(10 ppm C)
1
5.80
95.1
10.2
2
5.69
98.0
10.1
3
5.59
90.9
10.4
4
5.68
104
10.4
5
5.69
93.2
10.2
6
5.53
98.0
10.2
7
5.49
93.3
10.4
8
5.70
101
9.91
9
5.57
103
9.86
Mean
5.66
97.3
10.19
Stand. Dev.
0.12
4.50
0.20
RSD
2.13%
4.63%
2.0%
Accessories and Options
GE Autosampler
For high-volume laboratory applications,
the InnovOx can be used with the GE Au-
tosampler. It can handle up to 120 35-mL
sample tubes or up to 63 40-mL or 60-mL
vials in one protocol. An optional stirring
station creates a homogeneous mixture
for analysis during the sampling process.
Air Filter
The Air Filter is used to purify ambient air and pres-
surize it sufficiently for use as sparge and carrier
gas. It easily attaches to the back of the Lab model,
so the Analyzer can be used in place of external
carrier gas cylinders.
Sievers Certified Standards and TOC Vials
Sievers Certified Reference Materials are available for InnovOx
calibration and verification needs, as well as reliable and eco-
nomical 40-mL and 60-mL TOC vials.