7.3.4
The Signal Intensity is not Constant
Cause
Comment
Solution
Fiber optic cable has
shifted in position
Tape fiber optic cable to
bench
Lamp is not stable
Defective lamp
Replace the lamp
Sample is not
homogeneous
(Cuvette)
Mix sample and filter if
necessary to remove
particulate matter
Sample is not
homogeneous
(Integrating Sphere)
Move integrating sphere to
a different region of the
sample
Sample is decomposing
Replace sample
Cuvette is scratched
Always place cuvette in
sample holder with same
plane facing the source
Replace cuvette
Defect in Spectrometer
Contact Polychromix or its
representative
7.3.5
The Wavelength for a Peak has Shifted
Cause
Comment
Solution
The background spectrum
is not valid
Run a new background
Spectrum
Sample is not
homogeneous
Find another region of the
sample
Defect in Spectrometer
Contact Polychromix or its
local representative