Ocean Optics OOIIrrad-C User Manual
Page 57

5: Calibration
►
Procedure
1. Create one VIS calibration and label the file with the spectrometer’s serial number and _VIS.
Then, create one UV calibration and label the VIS calibration file with the spectrometer’s serial
number and _UV. The extension .cal is added automatically to both files.
2. On
the
Irradiance Calibration screen, select Spectrometer | Combine UV/VIS Calibrations
from the menu bar. The Combine UV/VIS Calibrations screen appears.
3. On
the
Combine UV/VIS Calibrations screen, select File | Open UV Calibration from the
menu bar. The name of the UV calibration file that you just selected displays.
4. On
the
Combine UV/VIS Calibrations screen, select File | Open VIS Calibration from the
menu bar. The name of the VIS calibration file that you just selected displays.
5. All other default values should be OK, so select Spectrometer | Combine UV/VIS from the
menu bar.
000-200000-070-0905