Spectrometer, Splice (splice) – Ocean Optics Ocean View Install User Manual

Page 100

Advertising
background image

5: Schematic View

92

000-20000-300-02-201312

Spectrometer

Enables you to input data to and output data from the connected spectrometer.

Location: Click the Device Manage button and select Rescan or Add Device to insert a new spectrometer
node on the schematic diagram.

Number of Inputs: Varies based on the spectrometer model.

Type of Inputs: Only output control nodes can be wired into the spectrometer node.

Type of Output: scalar or array

User Controls: Varies based on the spectrometer model.

Notes:

The icon for this node varies depending on the type of spectrometer connected. The serial number for the
device appears under the icon.

Right-click on this node to bring up a selection menu for additional input and output options.

Double-click on the Spectrometer node to display a list of device features and specifications.

Splice (Splice)

Combines multiple arrays into one larger array and outputs the combination.

Location: Advanced Math -->Array Math --> Splice

Number of Inputs: infinite

Type of Inputs: SubRange Node

Type of Output: array

User Controls: Starting Wavelength slider, Ending Wavelength slider, Full Range button, and Remove
Data Range button for each SubRange node; Apply and Add Range buttons

Notes:

A SubRange node must feed into the Splice node for each data array to be spliced together. To

keep the y-axis units consistent, the SubRange nodes should not get their data from different source types
(e.g., relative irradiance and absolute irradiance).

►

Procedure

1. Once all the SubRange nodes have been wired into the Splice node, double-click the Splice node

to open it and set the x-axis ranges for the different arrays. The slider bars allow easy selection of
starting and ending values for each array. The final combination is shown in the bottom graph.