4: Wizards

000-20000-310-02-201312

45

Quick View Minus Background

Select Quick View mode to take a Quick View and subtract the stored background spectra from each
spectrometer channel before OceanView displays it.

►

Procedure

1. Invoke the Quick View Minus Background wizard by clicking the Create new spectroscopy

application icon

.

2. Select either an active acquisition or start a new one.

3. Click Spectroscopy, then Quick View Minus Background, then Next.

4. Set Acquisition parameters.

5. Store background spectrum.

Absorbance

Absorbance spectra are a measure of how much light a sample absorbs. For most samples, absorbance
relates linearly to the concentration of the substance. OceanView uses an equation to determine the
concentration of a species in solution (illustrated below). The software uses this equation to evaluate each
pixel on the detector and produce the absorbance spectrum:

Where:

S = Sample intensity at wavelength

D = Background intensity at wavelength

R = Reference intensity at wavelength

The concentration of a species in a solution directly affects the absorbance of the solution. This
relationship, known as Beer’s Law, is expressed as:

Where:

A = Absorbance at wavelength

,

= Extinction coefficient of the absorbing species at wavelength

c = Concentration of the absorbing species and l is the optical path length of the absorption.