Refer to the following figure and table of pinouts while setting up your laser.
Laser Mechanical Drawing
Module Pinouts
Pin # Symbol
Description
1 NC
Not
connected
2
V set ENABLE Enables “LD SET” on Pin 8 when connected to Ground. If left open or set to 3 –
5 Volt, output power defaults to internally preset value.
3 T
SENS
4 T
SENS
1000 Ohm RTD sensor (with reference to Ground) to monitor module case.
5 GND
Ground
6
+ 5V
4.9 to 5.1 Volt; 1 Ampere
7
ENABLE
Tie to GND to disable laser output. Keep unconnected or apply 3 – 5 Volt to
enable laser output.
8
LD SET
Apply 0 to 1 Volt to control optical output power. Pin 2 must be grounded to
enable this option.
9 PD+
Photodiode
anode
10 PD-
Photodiode
cathode
4