Advanced instruments inc, 6 maintenance – Analytical Industries GPR-3100 Series Oxygen Purity Analyzer User Manual

Page 31

Advertising
background image

Advanced Instruments Inc.

6 Maintenance


GPR-3100 Rev 7/2004 6

1

Periodically, the oxygen sensor will require replacement. The operating life is determined by a number of

factors that are influenced by the user and therefore difficult to predict. Cleaning the electrical contacts

when replacing the sensor is the extent of the maintenance requirements of this analyzer.
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the

operator to service. Only trained personnel with the authorization of their supervisor should conduct

maintenance.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive liquid electrolyte that could

be harmful if touched or ingested, refer to the Material Safety Data Sheet contained in the Owner’s

Manual appendix. Avoid contact with any liquid or crystal type powder in or around the sensor or sensor

housing, as either could be a form of electrolyte. Leaking sensors should be disposed of in accordance

with local regulations.
Sensor Replacement Procedure
1. Using the 5/16 wrench supplied loosen but do not

remove the clamp bolt.

2. Rotate the upper section of the sensor housing 90º to

disengage from the clamp.

3. Remove the upper section by pulling it straight up and

place it on a smooth surface.

4. Remove the old oxygen sensor and dispose of it in

accordance with local regulations for disposing of

batteries.

5. Remove the new oxygen sensor from the shipping bag.
6. Remove the red label and the gold ribbon (shorting

deveice) from the PCB at the rear of the sensor.

7. Caution: Minimize the time the sensor is exposed to

ambient air.

8. Place the new sensor in the bottom section of the sensor housing with the PCB facing up.
9. Place the upper section of the sensor housing over the sensor.
10. Gently push the upper section downward and rotate 90º to engage the clamp.
11. Finger tighten the clamp bolt
12. Tighten the clamp bolt it one full turn with the 5/16 wrench to compressed the o-ring seal.
13. Sensor replacement is complete proceed to Zero and Span Calibration.

Advertising