Advanced instruments inc, 2 quality control certification – Analytical Industries GPR-3100 Series Oxygen Purity Analyzer User Manual

Page 4

Advertising
background image

Advanced Instruments Inc.


2 Quality Control Certification

Date:

Customer: Order No.:

Pass

Model

GPR-3100 High Purity % O2 Analyzer

Sensor

GPR-11-120-OP

Accessories

Owner’s Manual

CABL-1008 Power Cord Black 3 Conductor UL/CSA

HRWR-1021 Wrench 5/16 Combination

Configuration

Ranges: 0-100%; suppressed 50-100%, 80-100%, 90-100%

A-1146-E-50 PCB Assembly Micro-processor / Display

A-1147-E-50 PCB Assembly Power Supply / Interconnection

Software Rev: _______

Wetted parts: Stainless steel

Power: Specify: ( x ) 100/120 VAC, ( ) 220/240 VAC

Enclosure: ( ) Panel mount 10.8”W x 7.5”H

( ) Rack mount panel 19”W x 12”H

( x ) Wall mount 12”W x 12”H x 8”D

Test - Electronics

Factory default zero (without sensor)

Factory default span @ 20 uA

Alarm relays activate/deactivate to changes in O

2

concentration

Sensor failure alarm

Power failure alarm

Analog signal output 0-1V and 4-20mA

Range ID voltage output

Test - Gas Phase

Baseline drift <

± 1% FS 90-100% range over 24 hour period

Noise level <

± 1.0% FS

Span adjustment within 10-50% FS

Temperature controller (CNTL-_______) set at 85

°F

Final

Overall inspection for physical defects

Notes

1 of 1 analyzer due ASAP

GPR-3100 Rev 7/2004 2 1 of 1

Advertising