Range Mode Selection

Advance the cursor (*) to the “Sample” option as illustrated and press the green ENTER key to accept the selection.

MAIN MENU

* Sample
Span
Zero
Alarm
System
Standby

Standby

Auto Range
85⁰F 100Kpa


12/31/07 12:00:00


The following menu appears:

* SAMPLE

Auto Range
Manual Range
Bypass
Standby

Standby

Auto Range
85⁰F 100Kpa


12/31/07 12:00:00


The analyzer is equipped with five (5) standard measuring ranges (see specification) and provides users with a choice of
sampling modes. By accessing the MAIN MENU, users may select either the Auto Range or a fixed Manual Range) mode.

Note:

For calibration purposes, use of the Auto Range mode is recommended. However, the user can select a fixed Manual
Range (full scale) as dictated by the oxygen content of the calibration gas.