2 Quality Control Certification


Date:

Customer: Order No.:

Pass

Model:

GPR-2600 ppm Oxygen Analyzer S/N _____________________

Sensor:

( ) GPR-12-333 ppm Oxygen Sensor
( ) XLT-12-333 ppm Oxygen Sensor S/N _____________________

Accessories:

Owner’s Manual
CABL-1008 Power Cord
TOOL-1001 5/16” Combination Wrench

Configuration:

Ranges: 0-1 ppm, 0-10 ppm, 0-100 ppm, 0-1000 ppm

A-1146-40 PCB Assembly Main / Display Software V. ______
( ) A-1174-40 PCB Assy Power Supply / Interconnect, 4-20mA Range ID
( ) A-1174-40C PCB Assy Power Supply / Interconnect, 5x Relay Contacts Range ID

( ) Stainless steel sensor housing, manual flow control and bypass valves, ¼” compression
type fittings for sample inlet and vent
( ) Delete bypass valve from above (T and TO options)
( ) Sample, span, zero inlet solenoid valves

( ) Temperature controlled heater system 85°F specify: ( ) 110VAC ( ) 220VAC

Power: 100/120/220/250 VAC (universal without temperature controlled heater systems)

Enclosure: ( ) Std. panel mount (“T”) 7.5x10.8x12”; ( ) “TO” option 7.75x 7.75x12”
( ) Bezel for 19” rack mmount 19x12x12” option
( ) GPR-2600-W option general purpose wall mount 12x12x8”

Test

System start-up diagnostics satisfactory

Auto/manual range

Alarm relays activate/deactivate with changes in O

2

concentration

Alarm bypass

Analog outputs: Signal output 4-20mA

Range ID: ( ) 4-20mA or ( ) 5x relay contacts plus 1x common

Recovery from air to < 10 ppm in < 15 minutes

Baseline drift on zero gas <

± 2% FS over 24 hour period

Noise level <

± 1.0% FS

Span calibration gas value

Span adjustment within 10-50% FS

Peak to peak over/under shoot < 0.5% FS

Overall inspection for physical defects

Options

Label analyzer “Oxygen Service” in accordance with P-1507 Rev-1; see certificate next page.

Notes