Zero calibration, Span calibration, Considerations – Analytical Industries GPR-1500 A Series Trace PPM Oxygen Analyzer User Manual

Page 24

Advertising
background image

Advanced Instruments, Inc

24

The transmitter is equipped with a “Zero Calibration” feature. However, as described below, zero calibration
is recommended only when the application (or user) demands optimum accuracy of below 5% of the most
sensitive or lowest range available on the transmitter. For example, if the user requires analysis of a sample
gas below 0.5 PPM, a zero calibration is recommended.

Span calibration, in one of the forms described below, is necessary to adjust the transmitter sensitivity for
accurate measurements of oxygen contents in a sample gas. As a rule of thumb, zero calibration should be
carried out after span calibration.

Zero Calibration

Despite the absolute zero inherent in the electrochemical oxygen sensors, the reality is that analyzers may
display an oxygen reading even when sampling a zero gas (oxygen free gas) due to:

1. Contamination

or

questionable quality of the zero gas

2. Minor leakage in the sample line connections

3. Residual oxygen dissolved in the sensor’s electrolyte

4. Tolerances of the electronic components

The maximum “Zero Offset” of every analyzer is checked prior to shipment. However, due to the fact that the
factory sample system conditions differ from that of the user, no ZERO OFFSET adjustment is made at the
factory

Span Calibration

Involves periodically, see Intervals section below, checking and/or adjusting the electronics to the sensor’s
signal output at a given oxygen standard. The frequency of calibration varies with the application, e.g., the
degree of accuracy required by the application and the quality assurance protocol of the user. However, the
interval between span calibrations should not exceed three (3) months.

Note: Regardless of the oxygen concentration of the standard used, the span calibration process takes
approximately 10-15 minutes. However, the time required to bring a PPM analyzer back on-line after span
calibration can vary, see Online Recovery Time below.

Considerations

When it comes to the calibration of oxygen analyzers utilizing an electrochemical oxygen sensor,
circumstances vary widely from the ideal conditions that exist at the factory to a variety of differing
circumstances users encounter in the field. The following describes the most common factors and reasons
that influence the calibration procedures.

All electrochemical sensor based analyzers require periodic calibration, e.g. weekly intervals to a 3 month
maximum, to ensure accuracy and ascertain the integrity of the sensor

For optimum accuracy, calibrate the analyzer at or close to the temperature and pressure of the sample gas

The priority users place on getting or keeping an analyzer online is “the” most significant factor involved in
calibration and troubleshooting issues. The time it takes an analyzer to come down to a specific level after
exposure to high O2 concentrations or air is significantly different if a sensor is being installed than if the
sensor had been in-service at low oxygen levels for more than 1 week.

Advertising