The following displays appear:

000.00 %

PRESS UP OR DOWN

TO CHANGE VALUE

ENTER TO SAVE

MENU TO RETURN

>>>

0.55%

SPAN
CLAIBRATION
ENTER TO CALIBRATE
MENU TO ABORT

11. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the

underline cursor left to reach to the desired digit of the alarm value.

12. Repeat until the complete span value has been entered.

13. Press the ENTER key to accept SPAN CALIBRATION. After successful calibration, the analyzer will

display a message “PASSED CALIBRATION” and return to the Sample mode.

NOTE: The transmitter is allowed to accept calibration only when O2 reading is within 50% of the span gas
value. If the O2 reading is outside of this limit, by pressing ENTER to accept calibration will result in “FAILED
CALIBRATION” and the analyzer will return to the Sample mode without completing Span calibration.

PASSED CALIBRATION


OR




FAILED CALIBRATION

If the calibration is unsuccessful, return to the SAMPLING mode with span gas flowing through the
transmitter, make sure the reading stabilizes, reaches within 30-50% (see below) of the span gas value
(after factory default span setting) and repeat the calibration before concluding the equipment is defective.

Before disconnecting the span gas line and connecting the sample gas line (if the analyzer is not equipped
with a SPAN/SAMPLE vale option), flow the sample gas for 1-2 minutes to purge the air inside the sample
line, then disconnect the span gas line and replace it with the sample gas line.

Default Span

With factory default span, previous calibration data stored in the memory is removed and the sensitivity of
the analyzer is reset to the value based on the average output of the oxygen sensor at a specific oxygen
concentration. For example, with factory default settings, when a span gas is introduced, the micro-
processor will display oxygen reading within 30-50% of the span gas value, indicating that the sensor output
is within the specified limits. This feature allows the user to check the sensor’s signal output without
removing it from the sensor housing. To begin Factory Default span,

1. Access the MAIN MENU by pressing the MENU key.

2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.

3. Press the ENTER key to select the highlighted menu option.