Example: As illustrated by Graph A any error, play in the multi-turn span pot or the temperature compensation

circuit, during a span adjustment at 20.9% (air) of full scale range would be multiplied by a factor of 4.78

(100/20.9) if used for measurements of 95-100% oxygen concentrations. Conversely, an error during a span

adjustment at 100% of full scale range is reduced proportionately for measurements of lower oxygen

concentrations.
Recommendation: Calibrating with a span gas approximating 80% of the full scale range one or two ranges
higher than the full scale range of interest is recommended for 'optimum calibration accuracy'.