Analytical Industries GPR-1100 ATEX Portable Trace PPM Oxygen Analyzer User Manual

Page 16

Advertising
background image



16

Main Menu:
Access the MAIN MENU by pressing the MENU key:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO


Range Selection:
The GPR-1100 analyzer is equipped with five (5) standard measuring ranges (see specification) and provides users

with a choice of sampling modes. By accessing the MAIN MENU, users may select either the AUTO SAMPLING

(ranging) or MANUAL SAMPLING (to lock on a single range) mode.
Note: For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the
full scale MANUAL SAMPLE RANGE for calibration as dictated by the accuracy of the analysis required – for

example, a span gas with an 80 ppm oxygen concentration with the balance nitrogen would dictate the use of the

0-100 ppm full scale range for calibration and a 0-10 ppm measuring range.

Procedure - Auto Sampling:

1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor to highlight AUTO SAMPLE.
3. Press the ENTER key to select the highlighted menu option.
4. The display returns to the sampling mode:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN

DEFAULT ZERO




>>>

00.00 PPM

AUTO SAMPLING

10 PPM RANGE

24.5 C


The display will shift to the next higher range when the oxygen reading (actually the sensor’s signal output)

exceeds 99.9% of the upper limit of the current range. The display will shift to the next lower range when the

oxygen reading drops to 85% of the upper limit of the next lower range.

Advertising
This manual is related to the following products: