MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION



>>>

MANUAL RANGE

25%
10%
5%

1%

4. Advance the reverse shade cursor using the ARROW keys to highlight the desired MANUAL RANGE.
5. Press the ENTER key to select the highlighted menu option.
6. The following displays appears with the range selected and oxygen concentration of the sample gas:

MANUAL RANGE

25%
10%

5%

1%



>>>

3.3%

MANUAL SAMPLING

5 % RANGE


24.5 C

7. The display will not shift automatically. Instead, when the oxygen reading (actually the sensor’s signal output) exceeds

110% of the upper limit of the current range an OVER RANGE warning will be displayed.

8. Once the OVER RANGE warning appears the user must advance the analyzer to the next higher range via the menu and

keypad Press MENU, select MANUAL SAMPLING, press ENTER, select the appropriate MANUAL RANGE and press ENTER
again.







Start-Up is complete . . .