7 spare parts, 8 troubleshooting, Advanced instruments inc – Analytical Industries GPR-2000 ATEX Portable Oxygen Analyzer User Manual

Page 27

Advertising
background image

Advanced Instruments Inc.



27

Warning indicators:
An LED indicator located on the front panel will light continuously during the CHARGE cycle. A second LED indicator located on
the front panel provides a blinking 72 hour warning LOW BATT of the need to recharge the battery. Caution: Operating the
analyzer beyond this 72 hour warning may permanently damage the battery.

7 Spare Parts

Recommended spare parts for the GPR-2000 Series Portable Oxygen Analyzer:

Item No.

Description

GPR-11-32-RT

Oxygen

Sensor

GPR-11-21-RT

Oxygen

Sensor

XLT-11-24-RT

Oxygen

Sensor


Other spare parts:

Item No.

Description

A-1163

Battery Assembly (GPR-2000 only)

A-1157

Battery Assembly (GPR-2000P general purpose integral sampling pump)

A-1158

Battery Assembly (GPR-2000P intrinsically safe integral sampling pump)

FLTR-1002-1

Coalescing

Filter

FLTR-1004-1

Coalescing Filter replacement element

FLTR-1001

Filter

(GPR-2000P)

A-2420

Flow Through Adapter Assembly

A-2568

Housing Flow Through 1/8” SS Tube Fittings

MTR-1009

Meter Digital Panel LCD

A-1151-E-2

PCB Assembly Main / Display

PWRS-1002

Power Source Plug-in 9VDC 110V Battery Charger

PWRS-1003

Power Source Plug-in 9VDC 220V Battery Charger

A-2166-2

Pump

Assembly

A-2166-4

Pump Assembly Heavy Duty

A-2166-5

Pump Assembly Intrinsically Safe Design


8 Troubleshooting

Symptom

Possible Cause

Recommended Action

Reading does not reflect
expected values

Sensor was not calibrated at the pressure,
flow rate and temperature anticipated in the
sample gas stream

Recalibrate the analyzer

Oxygen reading drifts toward
zero

Indication sensor is nearing the end of its
useful life

Replace sensor, see section 6 Maintenance

Slow response time
Erratic oxygen reading

No oxygen reading

Liquid covering sensing membrane
Presence of interference gases
Unauthorized maintenance
Defective electrical connection
Sensor failure

Gently remove with damp lint free towel.
Consult factory

Consult factory
Replace sensor and/or consult factory

High oxygen reading

Inadequate control of flow rate
Abnormality in span gas

See Pressure and Flow
Qualify source

Advertising
This manual is related to the following products: