6 what to do if, 6what to do if – Xylem CG 842 User Manual
Page 36

What to do if...
79
6
What to do if...
Cause
Remedy
pH electrode:
– Not connected
– Connect electrode
– Air bubbles in front of the
diaphragm
– Remove air bubbles
– Air in the diaphragm
– Extract air or moisten
diaphragm
– Cable broken
– Replace electrode
– Gel electrolyte dried out
– Replace electrode
Cause
Remedy
pH electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
– Moisture in the plug
– Dry plug
– Electrolyte obsolete
– Replenish electrolyte or
replace electrode
– Electrode obsolete
– Replace electrode
– Electrode broken
– Replace electrode
Measuring instrument:
– Incorrect calibration
procedure
– Select correct
procedure
– Incorrect solution
temperature (without
temperature probe)
– Set up correct
temperature
– Socket damp
– Dry socket
Error message,
I /
Error message,