1. Remove probe guard.
2. IMPORTANT: Thoroughly dry the

sensor so that no water enters the
probe port when the sensor is
removed.

3. Insert the long end of the hex key

wrench into the small hole in the side
of the DO sensor. Turn the wrench
counter clockwise until the sensor is
released.

4. Pull the DO sensor out of the probe.

The DO sensor is keyed, or has a flat
side, so that it can not be removed or
inserted improperly.

5. Make sure that the inside of the connector and the o-ring of the sensor are clean and dry, with no

contaminants, such as grease, dirt, or hair. Insert the new sensor.

6. Use the hex key wrench to tighten the screw, making sure that the screw does not stick out on either

side of the DO sensor module. Also, if the hex screw was removed completely, make sure that it is
not cross-threaded when replaced.

7. The YSI 559 DO module is shipped with a dry, protective membrane. Before using for the first

time, remove the protective cap and replace it with a new one following the instructions for
Membrane Cap Installation.


ACCESSORIES/REPLACEMENT PARTS


The following parts and accessories are available from YSI or any YSI Authorized Dealer.

YSI Model Number

Description

5908

Membrane Kit, 1.25 mil PE, 6 caps and bottle of electrolyte solution

559

Replaceable DO Module

5238

Probe Reconditioning Kit, 10 sanding discs (400 grit) and sanding tool

5065

Form-Fitted Cover with Shoulder Strap

614

Ultra Clamp, C-Clamp Mount

4654 Tripod
5085

Hands Free Harness

5050

Small, Hard-sided Carrying Case, Foam-lined

5060

Small, Soft-sided Carrying Case, Precut Foam Interior

5080

Small, Hard-sided, Pelican Carrying Case, Precut Foam Interior


PROBE
GUARD

DO
SENSOR

CABLE

SET
SCREW

HEX KEY
WRENCH