Checking the probe zero, 3 membrane installation for the 5700-series probes, Membrane installation – YSI 58 User Manual

Page 12: Membrane installation for the 5700-series probes

Advertising
background image

Fill Membrane Cap

with Solution

Unscrew Cap

Remove

Stir Paddle

Screw Cap on

moderately tight

Replace

Stir Paddle

Figure 3


CHECKING THE PROBE ZERO


The probe zero is checked by immersing the probe in a sodium sulfite solution (0.08M or 3g
Na

2

SO

3

/300mL), or in water which has an inert gas bubbling through it (e.g. nitrogen, argon). The

meter should read less than 1% dissolved oxygen in either of these environments. If it does not,
change the membrane or clean the probe.

3.3.3 MEMBRANE INSTALLATION FOR THE 5700-SERIES PROBES

MEMBRANE INSTALLATION


All YSI 5700 Series Probes have similar sensors and should be cared for in the same way. They are
precision devices and require careful treatment if measurements of high accuracy are to be made.
Prepare the probe as described below.

1.

ADD DI OR DISTILLED

WATER

Figure

4


WARNING: All probes are shipped dry, you must follow these
instructions!

1.

Open the membrane kit and prepare the electrolyte solution.
Dissolve the KCl in the dropper bottle by filling it to the neck
with deionized or distilled water and shaking until the solids are
fully dissolved. After the KCL is dissolved, wait a few minutes
until the solution is free of bubbles.

2.

Remove the o-ring and the dry membrane from (after unscrewing the sensor guard.
Thoroughly rinse the sensor with electrolyte fluid.