SECTION 7 TROUBLESHOOTING

Symptom

Possible

Cause

Action

Ref

Low battery voltage

Replace batteries

2.3

Batteries installed wrong

Check battery polarity.

2.3

Instrument will not turn on

Meter requires service

Return system for service

8

Membrane is fouled or damaged Replace membrane

3.2

Probe anode is fouled or dark

Clean anode

5

Probe cathode is fouled

Resurface cathode

5

Instrument will not calibrate

System requires service

Return system for service

8

Instrument has recorded a shock Remove battery lid, wait 15

seconds for reset, and replace lid.

2

Batteries are low or damaged

Remove battery lid, wait 15
seconds for reset, and replace lid.

2

Instrument "locks up"

System requires service

Return system for service

8

Cal value is incorrect

Recalibrate w/correct value

4.3

Probe not in 100% water
saturated air during Cal
procedure

Moisten sponge & place in Cal
chamber w/ probe & Recal

4.3

Membrane fouled or damaged

Replace membrane

3.2

Probe anode is fouled or dark

Clean anode

5

Probe cathode is fouled

Resurface cathode

5

Dissolved Oxygen readings
are inaccurate

System requires service

Return system for service

8

LCD displays "LO BAT"

Batteries are low or damaged

Replace batteries

2.3

Allow sufficient time for
temperature and oxygen
stabilization

4.3

Check for water droplets on
membrane

3.2

Change membrane

3.2

Does not return to calibration
value

Clean probe (see instructions
under Electrode Cleaning)

5

Allow sufficient time for
temperature and oxygen
stabilization

4.4

Allow 60 sec for reading to
stabilize

Change membrane

3.2

Unstable readings


Clean probe (see instructions
under Electrode Cleaning)

5

YSI Incorporated

YSI 58 DO Meter

21