It is normal for a dark layer of silver chloride to cover the silver anode. After prolonged use it may
become necessary to clean the anode. Soak the probe in a 14% ammonium hydroxide solution for 2 to
3 minutes or overnight in a 3% ammonium hydroxide solution. Rinse with deionized water, recharge
with electrolyte, and install a new membrane.

5.3.2 5700-SERIES PROBE CARE

GOLD CATHODE


Inspect the gold cathode when changing the membrane; it should be bright and untarnished. If it is
untarnished, which can result from contact with certain gases, or plated with silver, which can result
from extended use with a loose or wrinkled membrane, clean it with a YSI 5680 Probe
Reconditioning Kit, or return it to an authorized YSI Repair Center. Never use chemicals or any
abrasive other than that supplied with the Reconditioning Kit.


SILVER ANODE


If the silver anode becomes contaminated, successful calibration can be prevented. Soak the probe
overnight in a 3% ammonia solution; rinse with deionized water, and follow the steps to install a new
membrane. If it is not possible to calibrate, return the probe to an authorized YSI Repair Center.

5.4 PROBE

STORAGE


For long term storage (4 weeks), remove the membrane, thoroughly rinse the sensor with deionized or
distilled water and install a new membrane. Store the sensor in a humid environment such as the
calibration chamber with the wet sponge inside. Do NOT store the probe dry.

5.4.1 5905 AND 5239 PROBE STORAGE


When the probe is not in use, store the probe in a BOD bottle containing at least 1 inch of water. For
long-term storage, remove the membrane cap, rinse the probe tip with deionized water, and install a
dry membrane cap (without electrolyte solution).

5.4.2 5700 SERIES PROBE STORAGE


Store the field probes in the plastic bottle provided. To keep the electrolyte from drying out, place a
small piece of moist towel or sponge in the bottle and insert the probe in one end. Store the BOD
probe in a BOD bottle containing at least an inch of water. Do not immerse the probe in water.