APPENDIX F

CALIBRATION VALUES TABLE

Calibration values for various atmospheric pressures and altitudes.
Note: This table is for your information only. It is not required for calibration.

Pressure Inches
of Hg

Pressure
mm Hg

Pressure
kPA

Altitude
in feet

Altitude
in meters

Calibration
Value in %

30.23 768 102.3 -276 -84 101
29.92 760 101.3 0

0

100

29.61 752 100.3 278 85

99

29.33

745 99.3 558 170 98

29.02

737 98.3 841 256 97

28.74

730 97.3 1126

343 96

28.43

722 96.3 1413

431 95

28.11

714 95.2 1703

519 94

27.83

707 94.2 1995

608 93

27.52

699 93.2 2290

698 92

27.24

692 92.2 2587

789 91

26.93

684 91.2 2887

880 90

26.61

676 90.2 3190

972 89

26.34 669 89.2 3496 1066 88
26.02 661 88.2 3804 1160 87
25.75 654 87.1 4115 1254 86
25.43 646 86.1 4430 1350 85
25.12 638 85.1 4747 1447 84
24.84 631 84.1 5067 1544 83
24.53 623 83.1 5391 1643 82
24.25 616 82.1 5717 1743 81
23.94 608 81.1 6047 1843 80
23.62 600 80.0 6381 1945 79
23.35 593 79.0 6717 2047 78
23.03 585 78.0 7058 2151 77
22.76 578 77.0 7401 2256 76
22.44 570 76.0 7749 2362 75
22.13 562 75.0 8100 2469 74
21.85 555 74.0 8455 2577 73
21.54 547 73.0 8815 2687 72
21.26 540 71.9 9178 2797 71
20.94 532 70.9 9545 2909 70
20.63 524 69.9 9917 3023 69
20.35 517 68.9 10293 3137 68