9

P

ROBLEM

S

OLVING

Trouble Shooting Guide

Use this quick reference table to correct symptoms you might be
experiencing.

Symptoms

Actions

Unable to
calibrate

Allow approximately 20 min. and up to 30

min. for probe to polarize when first turned
on

Change membrane

Clean probe (see instructions under Electrode

Cleaning)

Does not return
to calibration
value


Allow sufficient time for temperature and

oxygen stabilization

Check for water droplets on membrane

Change membrane

Clean probe (see instructions under Electrode

Cleaning)

Unstable readings

Allow sufficient time for temperature and

oxygen stabilization

Allow 60 sec for reading to stabilize

Change membrane

Clean probe (see instructions under Electrode

Cleaning)