13. Appendix A - Specifications

Materials: ABS, Stainless Steel, and other materials

Dimensions:

Height:

9.5 inches

(24.13 cm)

Thickness:

2.2 inches

(5.6 cm)

Width:

3.5 inches max.

(8.89 cm)

Weight:

1.7 pounds (w/ 10’ cable) (.77 kg)

Display:

2.3”W x 1.5”L

(5.8 cm W x 3.8 cm L)

Power: 6 AA-size Alkaline Batteries (included)

Approximately 100 hours operation from each new set of batteries

Automatic shutoff after 10 hours without a key press

Water Tightness: Meets or exceeds IP65 standards

Probe Operating Environment

Medium: fresh, sea, or polluted water and most other liquid solutions.

Temperature: -5 to +75 °C

Depth: 0 to 10, 0 to 25, 0 to 50, or 0 to 100 feet (depending on cable length)

Meter Ambient Operating/Storage Temperature: -5 to +45 °C

System Performance Specifications

Measurement Range Resolution

Accuracy

pH

0 to 14

0.01 unit

± 0.1 pH unit within ±10°C

of calibration temperature

or

± 0.2 pH unit within ±20°C
of calibration temperature

Conductivity

0 to 499.9

µS/cm

0 to 4999

µS/cm

0 to 49.99 mS/cm
0 to 200.0 mS/cm

0.1

µS/cm

1.0

µS/cm

0.01 mS/cm
0.1 mS/cm

± 0.5% FS

Salinity

0 to 80 ppt

0.1 ppt

± 2%, or ± 0.1 ppt

Temperature

-5 to +75 °C

0.1 °C

± 0.1°C ±1 LSD

Adjustable Conductivity Reference Temperature: 15°C to 25°C