Operating messages & error codes – YSI 910 COD Colorimeter User Manual

Page 13

Advertising
background image

13

englIsH

Operating Messages & error Codes

Operating messages

Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;

measurement is still possible.
A user calibrated method is indicated by an arrow while the test result is

displayed (see “Factory calibration reset“).

error codes

Light absorption too great. Reasons: e.g. dirty optics.
Calibration factor „out of range“
too much light reaching the detector.

too much light reaching the detector.
Battery capacity was too low during measurement. Change battery.

LR: Factory calibration incorrect / erased
LR: User calibration incorrect / erased
MR: Factory calibration incorrect / erased
MR: User calibration incorrect / erased

Battery Disposal

The 910 COD Colorimeter is powered by alkaline batteries which the user must

remove and dispose of when the batteries no longer power the instrument. Dis-

posal requirements vary by country and region, and users are expected to under-

stand and follow the battery disposal requirements for their specific locale.

Lo

E 10 / E 11

E 70

E 71

E 72

E 73

E 22

btLo

Time Cal

Store Date

RESULT

Cal

E27 / E28 / E29

E23 / E24 / E25

E 20 / E 21

Advertising