Troubleshooting, Specifications, Conversions – YSI EcoSense DO200A User Manual

Page 8: Recommended spare parts list

Advertising
background image

7

TROUBLESHOOTING

Main Display reads:

Possible Solutions:

“ovEr” or “undr”

Check membrane and electrolyte solution.

Clean anode and cathode.

Return product for service.

Secondary Display reads:

Possible Solutions:

“undr”

Heat the sample to above –6.0 °C

Return product for service.

“ovEr”

Cool sample to below 46.0 °C

Return product for service

SPECIFICATIONS

Display Range

Accuracy

Resolution

Dissolved O

2

(ppm or mg/L)

0 to 20.00 ppm (mg/L)

±2 % of the reading or ±2% air

saturation, whichever is greater

0.01 mg/L

Dissolved O

2

% air-sat

0 to 200.0 %

±2% of the reading or ±0.2 ppm,

whichever is greater.

0.1 %

Temperature °C

-6.0 to 46.0 °C

21 to 115 °F

±0.3 °C ±1 digit

0.1 °C

Pressure Compensation

600 to 1100 mBar (450 to 825 mmHg)

Salinity Compensation

From 0.0 to 40.0 ppt

ATC Probe

Thermistor, 10K, at 25C

Calibration Backup

Yes

Audio Feedback

Yes, on all keys

Power Source

One 9V battery

Operating Temperature

0 to 50C (32 to

122 °F)

Instrument Casing

Waterproof IP 67

Weight (with battery)

270 grams (.61 lbs)

Dimensions (L x W x D)

18.7 cm x 7.6 cm x 3.8 cm (7.3 in x 3 in x 1.5 in)

CONVERSIONS

To Convert:

Multiply by:

Inches of Hg to mBar

33.864

Inches of Hg to mmHg

25.4

mmHg to mBar

1.333

RECOMMENDED SPARE PARTS LIST

Item #606035REF

Revision A; July 2012

For the latest version of this manual, visit www.ysi.com

PART #

DESCRIPTION

200-4

4 meter (approx. 13 feet) probe and cable assembly

200-10

10 meter (approx. 33 feet) probe and cable assembly

200-BOD Self-stirring BOD lab probe and cable assembly with power supply

485

DO carrying case, soft sided

606036

DO carrying case, hard sided

5908

Membrane kit, 1.25 mil PE (605306), six cap membranes and KCl solution