Specifications, Recommended spare parts list – YSI EC300A User Manual

Page 10

Advertising
background image

8

SPECIFICATIONS

Display Range

Accuracy

Resolution

Conductivity,

Auto-ranging

0.0 to 499.9 uS/cm
500 to 4999 uS/cm
5.00 to 49.99 mS/cm
50.0 to 200.0 mS/cm

±1% of reading plus 2 uS/cm
±1% of reading plus 5 uS/cm

±1% of reading plus 0.05 uS/cm

±2.5% of reading plus 0.5 mS/cm

0.01 uS/cm

1 mS/cm

0.01 mS/cm

0.1 mS/cm

Salinity

0.0 to 70.0 ppt

0.2% Full Scale

0.1 ppt

Temperature

°C

-10.0 to 90 °C

±0.2 °C or ±0.4% Full Scale,

whichever is greater

0.1 °C

Reference Temperature

15.0 to 25.0

°C

Temperature Coefficient

0.0% to 4.0%

Cell Constant

5.00

±

0.50

TDS Constant Range

0.30 to 1.00

Power/Battery life

One 9V battery

Approximately 500 hour

Calibration Back-up

Yes

Audio Feedback

Yes, on all touch keys

Instrument Case

Waterproof when connector cap installed, IP 67

Operating Temp. Range

0 to 50

°C

Operating Relative Humidity Range

up to 95%

Temperature Probe

Thermistor, 10k / 25

°C

Dimensions (L x W x H)

18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in)

Weight (batteries included)

270 grams (.6 lb)

RECOMMENDED SPARE PARTS LIST

PART #

DESCRIPTION

300-1

1-meter probe and cable assembly

300-4

4-meter probe and cable assembly.

300-10

10-meter probe and cable assembly.

606043

Carrying case, hard sided.

485

Carrying case, soft sided.














Item #606042REF

Revision A, July 2012

For the latest version of this manual, visit www.ysi.com