Operation, Ongoing maintenance, Error displays and troubleshooting – YSI EC30A User Manual

Page 2

Advertising
background image

OPERATION

A. Power ON/OFF

Press the “On/Off” key. The unit will turn on and enter the “Measure Mode”. Press the key
again to turn off. The instrument has a 10 minute automatic shutoff if there is no key activity
unless it is submersed in solution.


B. Calibrate

1. Press “CAL” key to enter “Calibration” mode.
2. Immerse meter into 1000 µS conductivity solution.
3. The “CAL” icon will appear when the main display shows “1000”. Single Point calibration is

now complete. The “MEAS” icon will appear and the meter is ready to measure 0-2000 µS.
Rinse electrode.

4. Repeat Steps 2 and 3 for two/second point calibration by immersing the meter in 10,000 µS

(10 mS) conductivity solution and pressing the “Cal” key. Once completed, Two Point
calibration is successful and the meter is ready to measure 0-20 mS.

Note: For accuracy, it is recommended conductivity calibration is performed routinely with good
solutions and a clean electrode or anytime after replacing the electrode.


C. Measure

Dip the electrode into the test solution while in the “Measure Mode”. Give the meter a shake
while submersed to remove any air bubbles that may have occurred during submersion.


D. Hold Data

1.

When the conductivity is stable, press “HOLD” key once to lock the reading on the display.

2.

Press “HOLD” key again to unlock reading and the unit will return to “Measure Mode”.
The unit is now ready for another measurement.

ONGOING MAINTENANCE

A. Electrode Storage

1.

Electrodes attached to the instrument should be kept in a clean, dry environment. The
electrodes need to be kept clean.

B. Battery Disposal

This instrument is powered by alkaline batteries, which the user must remove and dispose of
when the batteries no longer power the instrument. Disposal requirements vary by country and
region, and users are expected to understand and follow the battery disposal requirements for
their specific locale.

ERROR DISPLAYS AND TROUBLESHOOTING

Conductivity
LCD
Display

ATC
Display

DISPLAY
Mode

Possible cause(s)
[Action(s)]

“OVER” "OVER"

Measure

a. Temperature outside 0 to 99.5°C range.
[Bring solution to a lower/higher temp.]
b. Bad sensor. [Replace electrode.]

"OVER"

0.0-
60.0°C

Measure

Conductivity value beyond 20 mS range .
[Use new solution, Recalibrate, or Replace
electrode.]

"OVER"

60.0-
99.5°C

Measure

Temperature >60.0°C, over the temperature
compensation range.
[Bring solution to a lower temperature or replace
electrode.]

“OVER” “udr” Measure

Temperature <0.0°C, under the temperature
compensation range.

Advertising