NitraVis 70x IQ TS

Technical data

8 - 1

ba75950e02

11/2012

8

Technical data

8.1

Measurement characteristics

Measuring principle

Spectrophotometric absorption measurement; integrated micropro-

cessor electronics, shielded 2-wire connection for

power and data transmission.

Wavelength range

200 - 720 nm

Measuring gap (optical

layer thickness)

Nitrate measurement,

1 mm gap width

Nitrate measurement,

5 mm gap width

TSS measurement,

1 mm gap width

TSS measurement,

5 mm gap width

The quoted measuring ranges are nominal measuring ranges that are

theoretically possible. In practice, real measuring ranges exist that are

given by the limits of photometric determination. The limits are signifi-

NitraVis 701 IQ TS

1 mm

NitraVis 705 IQ TS

5 mm

Measuring

location

Measuring

mode

Measuring range

Resolution

Inlet

NO3

0.0 ... 300.0 mg/L

0.1 mg/L

NO3-N

0.00 ... 60.00 mg/L

0.01 mg/L

Activation

NO3

0.0 ... 300.0 mg/L

0.1 mg/L

NO3-N

0.00 ... 60.00 mg/L

0.01 mg/L

Outlet

NO3

0.0 ... 750.0 mg/L

0.1 mg/L

NO3-N

0.0 ... 150.0 mg/l

0.1 mg/L

Measuring

location

Measuring

mode

Measuring range

Resolution

Outlet

NO3

0.0 ... 250.0 mg/L

0.1 mg/L

NO3-N

0.00 ... 50.00 mg/L

0.01 mg/L

Measuring

location

Measuring

mode

Measuring range

Resolution

Inlet

TSS

0.00 ... 15.00 g/L

0.01 g/L

Activation

TSS

0.00 ... 20.00 g/L

0.01 g/L

Measuring

location

Measuring

mode

Measuring range

Resolution

Outlet

TSS

0.0 ... 900.0 mg/L

0.1 mg/L