YSI IQ S ENSOR N ET NitraVis 70x IQ TS User Manual

Page 20

Advertising
background image

Commissioning

NitraVis 70x IQ TS

3 - 10

ba75950e02

11/2012

Cal - # raw value 1

Value pairs of the user calibration (see section 4.3.2).

Note: Leave those values unchanged during the initial commission-

ing.

Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Zero adjustment

Factory

User

Setting of which zero adjustment is

used in the sensor.

Factory zero adjustment or

Zero adjustment carried out last

Signal smoothing

On

Off

Switching on/off the signal smoothing.

For more details, see

Response time

t90

.

Response time t90

10 ... 60 min

Response time of the signal smoothing.

Depending on the sample matrix, mea-

sured values can vary more or less

strongly (e.g. due to foreign bodies or

air bubbles).

The signal filter (here, the

Response

time t90

)

reduces the limits of variation

of the measured values.

The

Response time t90

specifies the

time after which 90 % of a signal

change are displayed.

This menu item is only available with:

Signal smoothing

=

On

The specified time is used as a guide

value for low (10 min) to high (60 min)

signal smoothing and can deviate from

the value that was set.

Cleaning mode

On

Off

Switching on/off the automatic, time-

controlled cleaning function.

Menu item

Settings

Explanations