2 status info, 2 status info -5 – YSI IQ S ENSOR N ET NitraVis 70x IQ TS User Manual

Page 55

Advertising
background image

NitraVis 70x IQ TS

Indexes

9 - 5

ba75950e02

11/2012

9.2

Status info

The status info is a coded piece of information on the current status of

a sensor. Each sensor sends this status info to the controller. The sta-

tus info of sensors consists of 32 bits, each of which can have the value

0 or 1.

Status info,

general structure

The bits 0 - 15 are reserved for general information.

The bits 16 - 21 are reserved for internal service information.
You obtain the status info:

via a manual query in the

Settings

/

Service

/

List of all components

menu (see system operating manual)

by an automated query
– of a superordinate process control (e. g. when connected to the

Profibus)

– of the IQ Data Server (see operating manual of the

IQ S

ENSOR

N

ET

software pack)

The evaluation of the status info, e.g. in the case of an automated

query, has to be made individually for each bit.

Status info

NitraVis 70x IQ TS

0 1 2 3 4 5 6 7

8 9 10 11 12 13 14 15

1 0 0 0 0 0 0 0

0 0 0 0 0 0 0 0 (general)

0 0 0 0 0 0 0 0

0 0 0 0 0 0 0 0 (internal)

16 17 18 19 20 21 22 23

24 25 26 27 28 29 30 31

Status bit

Description

Bit 0

Component hardware defective

Bit 1

Component hardware defective

xxx

Bit 2

Optical measuring range exceeded

Bit 3-31

-

Advertising