2 fields of application, 3 measuring principle of the uv-vis sensor – YSI IQ S ENSOR N ET NitraVis 70x IQ TS User Manual

Page 6

Advertising
background image

Overview

NitraVis 70x IQ TS

1 - 2

ba75950e02

11/2012

1.2

Fields of application

The range of application comprises control, feedback control and mon-

itoring.

Measuring location

Ultrasound cleaning

system

The ultrasound cleaning system ensures low maintenance and long-

term reliable measurement operation.

The ultrasound source integrated in the sensor excites the measure-

ment windows to produce vibrations in the ultrasound range. The

movement of the surface as a result of this prevents the growth of

pollution in almost all cases and, thus, ensures reliable measured

values during continuous operation.

Compressed air cleaning

(option)

The sensor is designed to be equipped with an additional compressed

air-driven cleaning system, which supports the ultrasound cleaning

system as necessary.

1.3

Measuring principle of the UV-VIS sensor

The UV-VIS sensor operates according to the measuring principle of

UV-VIS spectrometry.
Any substance contained in a measuring solution attenuates a light

beam that penetrates the measuring solution. The attenuation (absor-

bance) of the light beam is measured for different wavelengths. The

wavelength dependent distribution produces the absorbance spec-

trum.
The wavelength range where an absorbance occurs is characteristic of

the substance. The extent of the absorbance depends on the amount

of the substance.
The influence of typical interference matrix variants occurring at munic-

ipal waste water treatment plants can be eliminated by including the

entire absorbance spectrum.
With this measuring principle, measuring does not take place continu-

ously but at intervals.

NitraVis 701 IQ TS

Aeration tank or inlet

NitraVis 705 IQ TS

Outlet

Advertising