4

Floor

Ceiling

Caution:

Follow these instructions to

avoid failure.

Danger:

Wear eye protection when

installing or removing flowmeter.

1. Misalignment will damage the meter!

Flowmeter must be installed in an
exact vertical plane to ensure accu-
racy. Be certain of proper plumbing
alignments. Misalignment may cause
the o-ring seals to leak. The
meterbody material can be damaged
by UV rays. Do not install in direct
sunlight.

2. Pipe dope and glue will damage
the meter!

Use only Teflon® tape on the threaded
adapters. Polysulfone meter body and
fittings cannot tolerate PVC Glue
and/or pipe dope. Even fumes can
cause severe damage. If you are
installing your flowmeter to a glued
pipe configuration, install the flowmeter
after all glued fittings are dried and
lines are purged of all fumes. Never
hold the meter body with pliers or like
tools. Union nuts should be hand tightened only. DO NOT OVER-TIGHTEN!

3. Vibration and heavy loads will damage the meter!

Wall, floor and ceiling mounts and supports must be carefully aligned with the
meter body and sturdy enough to support the plumbing and prevent vibration.
Never allow the flowmeter to support the weight of related piping.

4. Solenoid valves will damage the meter!

Avoid a system that will impose a sudden burst of flow to the meter. Such a
burst will cause the float to impact the float stop with destructive force. Solenoid
valves, or other quick opening valves cannot be used unless meter is protected
against sudden bursts of flow.

5. High pressures and temperatures will damage the meter!

The maximum acceptable temperature and pressure is interdependent. The
maximum acceptable working pressure is dependant on the actual fluid tempera-
ture. The maximum acceptable fluid temperature is dependant on the actual
working pressure. (see Temperature Vs. Pressure chart).

Installation Guideline

Blue-White

Industries, Ltd.

R

Pressure

Temperature

and

Pressure and temperature limits are inversely proportional. At the maximum
suggested pressure the temperature should approach 70°F / 21.1°C; at the
maximum suggested temperature the pressure should approach zero psi. We
cannot guarantee our flowmeters will not be damaged either at or below the
suggested limits simply because of many factors which influence meter integrity;
stress resulting from meter misalignment, damage due to excessive vibration
and/or deterioration caused by contact with certain chemicals as well as direct
sunlight. These situations and others tend to reduce the strength of the materials
from which the meters are manufactured.

Flowmeters are tested and calibrated for water or air only.

Although meters may be suitable for other chemicals, Blue-White
cannot guarantee their suitability. It is the responsibility of the user
to determine the suitability of the flowmeter in their application.

Application Note

Temperature

PSIg / BAR

Temperature vs. Pressure

90°F / 32.2°C

70°F / 21.1°C

0 / 0

30 / 2.1

190°F / 87.7°C

170°F / 76.6°C

150°F / 65.6°C

130°F / 54.4°C

110°F / 43.3°C

212°F / 100°C

60 / 4.1

90 / 6.2

120 / 8.3

150 / 10.3

Maintenance

The “Exploded View” drawing illustrates assembly of the F-450 series meter. If
your flowmeter needs to be cleaned refer to this drawing when reassembling the
unit. The tapered tube may be cleaned with a soft bottle brush. Use a MILD soap
and water solution for cleaning purposes. Hard water deposits can be removed
with a 5% acetic acid solution (vinegar). Note the floats “up” position.