Xylem EX100/200-Series ELECTROMAGNETIC FLOW SENSORS User Manual

Page 10

Advertising
background image

Grounding Guidelines:

For best results, use a good quality earth ground, such as
metallic water piping or a driven ground, to ensure a good
connection to earth ground and good noise suppression.

If the flow sensor is installed in metallic piping, for optimum
connection clamp wire to the piping a short distance to one
side of the flow sensor using an electrical grounding clamp.
Connect the wire to the earth ground and to one of the housing
screws.

For Non-Metallic Pipe:

Connect one to the housing screws by

wire to a good earth ground, such as metalic water piping or a
rod driven into the ground.

EX meters are usually unaffected by moderate levels of electri-
cal noise. In some applications performance may be improved
by taking the following steps:

• Use shielded twisted pair cable (Belden 8723 or

equivalent above ground or Alpha 35482 or

equivalent burial).

• Clamp a ferrite bead (Steward 28A2029-OAO or

equivalent) on meter signal/power wire within 3/4”

of the meter strain relief (tape or tie wrap in place if

necessary). See diagram below.

• IMPORTANT - Connect the cable shield ground

wire to ground, ONLY at power supply end of cable.

General Electrical Guidelines:

• Whenever possible avoid running control cables in

the same conduit with or bundled with AC power.

• Using shielded cable, be sure to connect shield to

ground at power supply end of the cable. Do not
connect other end of shield.

• Avoid routing flow sensor cables in close proximity to

a variable frequency drive.

• Recommended power and output wiring is shielded

twisted pair 18-22 AWG control cable.

• Recommended voltage is 12-24 Vdc. Note that

unregulated power supplies can vary from nameplate
voltage by a considerable amount, especially with AC
line voltage fluctuation. Therefore 24V power supplies
must be regulated.

See the Connections diagrams on the following pages for the
appropriate terminals.

Power:

A 12 - 24 Vdc power supply capable of at least 250

mA current output is needed.

Output:

The standard output is pulses with flow in the forward

direction (Reverse flow output is optional [-15] ). Electrically
it is an open collector opto isolated switch. The output is not
internally powered.

Note:

This output is limited to 6 mA at 30 Vdc maximum.

Page 8

ELECTRICAL CONNECTIONS

GROUNDING DIAGRAM

Housing Screw

Earth

Ground

Grounding Clamp

PLACE FERRITE BEAD HERE