Mark First Transducer Location (A)
Mark Second Transducer Location (B)
Measure Separation Distance
TRANSIT TIME V-MOUNT MODE
Centerline
Transit Time V mounting method:
When the meter is operated in the Transit Time method, marking the second transducer
location is required.
The V Mount mode requires careful separation distance measurements. Inaccurate
placement of the transducers may result in insufficient signal strength and poor
measurement accuracy.
Draw a straight centerline, parallel to the center line of the pipe, from the first
transducer location mark (A) through to the second transducer location (B). Place a mark
on the centerline at the correct transducer separation distance.
If the fluid contains little or no particles (up to 10% maximum), operate the meter using
the Transit Time method to obtain the best accuracy.
V-MOUNT MODE
6.7
Transit Time
Transducer
Installation
V Mount Mode
Ultrasonic Flowmeter