Cleaning the electrodes – Xylem CellOx 325 Dissolved oxygen sensor User Manual

Page 10

Advertising
background image

Maintenance, cleaning, replacement

CellOx 325

26

5.4

Cleaning the electrodes

Caution

Before starting to work with the sensor, please note the G

ENERAL

MAINTENANCE

INSTRUCTIONS

on page 23.

General

information

Cleaning is only required in cases of slopes that are too small or too large
(sensor cannot be calibrated) that cannot be resolved by changing the mem-
brane head and electrolyte solution.

Cleaning the electrodes

Unscrew the membrane
head.

Caution: Electrolyte solution!
For disposal of the membrane
head and electrolyte solution,
see section 5.6.

Rinse the sensor head with
deionized water.

Using the rough side of the
wet SF 300 polishing strip,
polish off any contamination
from the gold working elec-
trode using light pressure.

Caution: Do not use any con-
ventional sandpaper or glass-
fiber brushes.

Rinse the sensor head with
deionized water.

Wipe the counter electrode
with a lint-free paper towel
and carefully remove any
loose white deposits.

Immerse the sensor head in-
cluding the counter electrode
in RL/G cleaning solution.

Allow to react for 1 to 3 min-
utes.

SF 300

RL/G