2 input status, 1 valid command(s), 2 format – Xylem CL500 ONLINE CHLORINE ANALYZERS MODBUS User Manual
Page 9: Valid addresses, 4 definitions
Advertising

4.2 Input Status
These single-bit values are readable from the master. The data will be returned with the lowest
addressed input status in the LSB of the data. Unused bits in the data will be set to 0.
4.2.1 Valid Command(s)
Code Name
Broadcast
0x02 Read
Input
Status
No
4.2.2 Format
16-bit word format
MSB
LSB
Bit 15
Bit 14
Bit 13
Bit 12
Bit 11
Bit 10
Bit 9
Bit 8
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
4.2.3.Valid Addresses
10001 – 10XXX
4.2.4 Definitions
Address Function Default
10001 Instrument
error
False
10002
Instrument error with alarm (levels 1,2and 3)
False
10003
Alarm 1 active
False
10004
Alarm 2 active
False
CL500 Modbus (3/09)
Rev. 1.3
7