Storage and maintenance

Electrodes should be stored at a temperature between
0°C and 40°C. Depending on the storage conditions
(temperature and air humidity) the irrigation fluid in the cap
may dry out prematurely. pNa electrodes stored dry must be
irrigated for minimum 24 hour in sodium chlorid solution
0.1 mol/l before measurement. Crystals in the electrolyte
area can be dissolved by warming the electrode in a water
bath. The electrolyte solution should then be replaced.

The correct potassium chlorid concentration must be
observed.

Quality

Every electrode must fulfill the strict quality requirements of
final testing. The working life depends mainly on the usage
conditions. Extreme measuring conditions as for example
high or frequently fluctuating temperatures, strong acids and
caustic solutions, protein and heavily contaminated
solutions, sulfides, bromides and iodides shorten the
electrode’s working life. Hydrofluoric acid, sodium hydroxide
and hot phosphoric acid attack glass.


Technical details Na 61

pNa range:

0 ... 6

Temperature range:

-10 ... +80°C

Length:

170 mm

Sensor:

Na sensitive glass

Diaphragm: platinum

Reference systeme:

Silamid

®

Material:

glass, d = 12 mm

Zero point:

pNa = 2.0

Electrolyt:

3 M KCl

Irrigation fluid:

0.1 M NaCl

Connection cable:

e.g. L 1 A

Temperature sensor:

no


Order number

Combination electrode Na 61

285100026

Connection cable

L1A

285122456

Electrolyte 3M KCl

L 3008 285138505

Cleaning solution

L 510

285138538


Further information

Further information can be found in the labratory catalogue
of SI Analytics GmbH.