Xylem CR500 CIRCULAR CHART RECORDERS User Manual
Page 28

Global Water
800-876-1172
• globalw.com
- 28 -
5-1 Input Range Table:
Type Range
Color
Code Polarity
J
-130° C / 760° C
(-202° F / 1400° F)
White
Red
+
-
K
-130° C / 1370° C
(-202° F / 2498° F)
Yellow
Red
+
-
T
-200° C / 400° C
(-328° F / 752° F)
Blue
Red
+
-
B
100° C / 1824° C
(212° F / 3315° F)
Gray
Red
+
-
R
0° C / 1650° C
(32° F / 3002° F)
Black
Red
+
-
S
0° C / 1649° C
(32° F / 3000° F)
Black
Red
+
-
PT100
(DIN)
-210° C / 440° C
(-346° F / 824° F)
mA
4-20 Ma
Table 5-1 Input Characteristics
Lower Chart Range Selection:
Programmable from - 999 to 9999 Units
Upper Chart Range Selection:
Programmable from - 999 to 9999 Units
Table 5-2 Upper/Lower Chart Ranges
Note: Recommended minimum chart span for thermocouple inputs is 100