5-1 Input Range Table:

Type Range

Color
Code Polarity

J

-130° C / 760° C

(-202° F / 1400° F)

White
Red

+

-

K

-130° C / 1370° C

(-202° F / 2498° F)

Yellow
Red

+

-

T

-200° C / 400° C

(-328° F / 752° F)

Blue
Red

+

-

B

100° C / 1824° C

(212° F / 3315° F)

Gray
Red

+

-

R

0° C / 1650° C

(32° F / 3002° F)

Black
Red

+

-

S

0° C / 1649° C

(32° F / 3000° F)

Black
Red

+

-

PT100

(DIN)

-210° C / 440° C

(-346° F / 824° F)

mA

4-20 Ma




Table 5-1 Input Characteristics

Lower Chart Range Selection:

Programmable from - 999 to 9999 Units


Upper Chart Range Selection:

Programmable from - 999 to 9999 Units


Table 5-2 Upper/Lower Chart Ranges

Note: Recommended minimum chart span for thermocouple inputs is 100