Xylem Flow Converter 713 User Manual

Page 16

Advertising
background image

16

previously keyed in quantities. The flow can only be set

in whole numbered m

3

/h or l/s. The highest Q(h) gives

the measuring field.

3.9.1 Optional formula enter exponent

The exponent in the flow formula (Q(h)=K*h

x

) is keyed

in. Q is the flow in m

3

/h, h presents the level in meters,

K is a factor between 1 and 9999 and x is the exponent

which is between 1.000 and .500, K and x is unitless.

Key in with the arrow keys. Double arrow runs the value

fast with 0.1 steps. Single arrow runs the value slowly

with 0.001 steps.

3.9.2 Optional formula enter factor

The factor K in the flow formula (Q(h)=K*h

x

) is keyed in.

Key in with the arrow keys. Double arrow runs the value

fast with 100 steps. Single arrow runs the value slowly

with 1 steps. The factor can be adjusted in the area 1

- 99999.

4.0

Programming of digital outputs

4.1

Enter digital output

Configuring of the 5 digital outputs. Select with the ar-

row keys which digital output to program, confirm with

ENTER.

Digital output 5 (DO5) is standard an output for an exter-

nal counter. DO5 can be ordered as an option for a relay

output.

4.2

Select function for DO

For digital output 1-4 select with arrow keys between

8 various functions: counter output, sampler, flow>0%,

flow high, flow low, alarm 24 hour volume, alarm 1 hour

volume and alarm sensor error. The choice is confirmed

with ENTER.

Counter output:

After a programmed number of m

3

,

the output is activated for an

external counter.

Sampler:

After a programmed number of m

3

,

the output is activated for start up

of an externally connected

sampler, or to a possibly connected

chemical dosing.

Flow>0%:

Signal to indicate that flow is

greater than 0, is applied i.e. when

measuring emergency stormflow.

Alarm flow high:

Activated if flow exceeds an

adjusted value.

Alarm flow low:

Activated if flow drops below an

adjusted value.

4 hour volume:

Activated is 4hour volume alarm

exceeds a programmed value.

Hour volume:

Activated if hourly volume exceeds

a programmed value.

Sensor error:

Activated at sensor error.

Out of action:

Is chosen when the output is not

applied.

4.3

Alarm flow high

The limit for flow high is keyed in with the arrow keys.

Double arrow runs the value fast with 10cm steps, single

arrow runs the value slowly with 0,1 cm steps. The area

is 0.0 to (maximum flow+10%).

4.4

Alarm flow low

The limit for flow low is keyed in with the arrow keys.

The area is 0.0 to (maximum flow+10%).

4.5

Alarm 24 HR volume

The limit for max. 4 hour volume is keyed in with the

arrow keys.

4.6

Alarm 1 HR volume

The limit for max. hourly volume is keyed in with the ar-

row keys.

4.7

Enter signal delay

The time a limit for an alarm can be exceeded before a

DO is activated, is set. The format is as follows - hours:

minutes. The maximum delay is 99 hours and 59 min-

utes. The delay is working for chamge from not active to

active mode as well as change from active to not active

mode.

4.8

Enter volume between pulses

The volume in m

3

that passes between each time a pulse is

sent to a DO is keyed in.

4.9

Digital output ON-time

In this menu the ON-time for the DO is keyed in. Key in

with arrow keys. The area is (0,1-30 sec).

4.10 Digital output NC/NO

In this menu the relay function for digital output is keyed

in, as Normally Open (NO) or Normally Closed (NC).

Select with arrow keys, confirm with ENTER.

5.0

Programming of analog output

In this menu the mA-output of the flow converter is

adjusted.

5.1

Analog output 0-20 / 4-20 mA

Here you select between mA-output 0-0 or 4-0 mA.

Select with arrow keys, confirm with ENTER. The output

follows the flow at max flow, keyed in menu 2.5 gener-

ated 0mA.