Quick Guide for Oxix
Calibration
Version:
04-2012 KJ
EN 5.4 Oxix Quick guide for calibration 1204.doc
3/7
Zero Point Calibration
A zero point check or calibration is conducted as follows:
•
The sensor is set for a zero point solution. The current measurement is shown in line 2
(0.10mg/l in this example).
•
Set the desired measuring value on the first line using the up and down arrow keys.
•
Save the current measurement setup by pressing OK.
The value is saved in the converter and will correct the measurement value of the sensor.
See section Guide to Oxygen-free Solution for Checking Oxix D.O. Sensor Zero Point.