Zero point calibration – Xylem Oxix Calibration User Manual

Page 3

Advertising
background image


Item

Oxix

Name:

FW 846003-001
Release date: 12/07 2011
S/N at release date: 97060

Quick Guide for Oxix
Calibration

Version:

04-2012 KJ

EN 5.4 Oxix Quick guide for calibration 1204.doc

3/7

Zero Point Calibration

A zero point check or calibration is conducted as follows:

•

The sensor is set for a zero point solution. The current measurement is shown in line 2

(0.10mg/l in this example).

•

Set the desired measuring value on the first line using the up and down arrow keys.

•

Save the current measurement setup by pressing OK.


The value is saved in the converter and will correct the measurement value of the sensor.
See section Guide to Oxygen-free Solution for Checking Oxix D.O. Sensor Zero Point.